Csi wirion recall
WebNov 25, 2024 · CSI recalls Wirion embolic protection device. 25th November 2024. 3322. Cardiovascular Systems has initiated a voluntary recall of unused Wirion embolic protection systems due to complaints of filter breakage during retrieval. Wirion is a distal embolic protection filter used to capture thrombus and debris that can be associated with all types ... WebFeb 8, 2024 · Cardiovascular Systems, Inc. Recall. On November 24, 2024, CSI announced the WIRION Embolic Protection System voluntary recall. The company initiated the recall of all unused devices “due to complaints of filter breakage during withdrawal.”. The company recalled all customer inventory of the WIRION Embolic Protection Device (Model WRN-D6).
Csi wirion recall
Did you know?
Web2 days ago · This product is distributed by CSI. Manufactured By. WIRION ® Peripheral Embolic Protection System. Indications Statement: WIRION® is indicated for use as an embolic protection system to contain and remove embolic material (thrombus/debris) while performing atherectomy in calcified lesions of the lower extremities. The diameter of the … WebNov 24, 2024 · The company plans to voluntarily recall all WIRION units currently in customer inventory. In total, 697 devices were distributed in the U.S. between March 22, …
The WIRION® system is used to hold and remove debris or blood clots from the lower limbs that can be associated with atherectomyExternal Link Disclaimerprocedures. See more Customers in the U.S. with questions about this recall should contact Cardiovascular Systems Inc. by phone at 651-259-2800. See more Cardiovascular Systems Inc. is recalling the WIRION system due to complaints of filter breakage during retrieval. Under certain … See more On November 22, 2024, Cardiovascular Systems Inc. sent an urgent notification recall letter to customers instructing them: 1. Remove the device from distribution and return the device to Cardiovascular Systems Inc. See more WebAug 6, 2024 · presentation regarding CSI’sgrowth, future financial measurements, product development and introductions (including the timing thereof), clinical trials (including the timing thereof), international expansion, the timing of manufacturing transfer of the WIRION system, product benefits, and market opportunities, are forward-looking statements.
WebCSI representative or call customer service at 1-877-274-0901. PACKAGE INCLUDES: (1) Delivery Catheter with Pre-Loaded Filter, (1) Retrieval Catheter (1) Syringe with Flushing Tip (5 cc) WIRION® is indicated for use as an embolic protection system to contain and remove embolic material (thrombus/debris) while performing atherectomy WebNov 24, 2024 · The company plans to voluntarily recall all WIRION units currently in customer inventory. In total, 697 devices were distributed in the U.S. between March 22, …
WebAug 5, 2024 · The Wirion device is a distal embolic protection filter that is used to capture debris that can be associated with all types of peripheral vascular interventions. CSI plans to commercialize the Wirion system in the United States after the transfer of manufacturing from Gardia Medical, which is expected to be completed after a 12- to 15-month ...
WebMar 17, 2024 · For example, statements in this press release regarding the benefits of the WIRION system; the commercialization of the WIRION system; and CSI’s … chinuku chinuku andelatho lyricsWebMar 17, 2024 · Device offers a streamlined approach for lower extremity embolic protection Cardiovascular Systems, Inc. (CSI ® ) (NASDAQ: CSII), a medical device company … granta edupack student instructionWebNov 26, 2024 · Print. Cardiovascular Systems Inc. announced a voluntary recall of its Wirion device on Wednesday afternoon. The company has told health care facilities to … grant a gift vinyl wrapWebNov 26, 2024 · CSI expects to introduce the acquired catheters as its new ViperCross line in the first half of its fiscal year, ending on 30 June 2024. Price Action: CSII shares closed … chinuku thadiki song lyrics teluguchinu limited rugger 65WebNov 24, 2024 · The company plans to voluntarily recall all WIRION units currently in customer inventory. In total, 697 devices were distributed in the U.S. between March 22, 2024 and November 15, 2024. All lots ... grant a gift autism foundation walkWebNov 24, 2024 · Cardiovascular Systems, Inc. (CSI®) (NASDAQ:CSII) has initiated a voluntary recall of unused WIRION Embolic Protection Systems (WIRION) due to complaints of filter breakage during retrieval. granta great shelford